Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
317
Operating Manual 1176.7578.02 ─ 09
[SENSe:]FREQuency:CENTer:STEP:AUTO
<State>
This command links the step width to the current standard (ON) or sets the step width
entered using the
FREQ:CENT:STEP
command (OFF) (see
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
FREQ:CENT:STEP:AUTO ON
Activates the coupling of the step size to the span.
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines the frequency offset.
Parameters:
<Offset>
Range:
-100 GHz to 100 GHz
*RST:
0 Hz
Default unit: Hz
Example:
FREQ:OFFS 1GHZ
Manual operation:
See
[SENSe:]SWEep:COUNt[:VALue]
<SweepCount>
This command sets the statistics count. Entering 0 as a parameter activates "Auto"
mode. Entering a number greater than 0 activates "Manual" mode and sets the statis-
tics count to the corresponding number.
For more information see
●
Setting parameters:
<SweepCount>
numeric value
Range:
0 to 32767
*RST:
0
Default unit: NONE
Example:
INIT:CONT ON
Activates continuous sweep mode.
SWE:COUN 0
Records the I/Q data continuously and uses a sliding window
length for averaging of 10.
INIT:CONT OFF
Activates single sweep mode
SWE 5
Records I/Q data until 5 evaluations have finished.
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...