Operating Manual 1176.7578.02 ─ 09
Example:
SENS:ADJ:CONF:HYST:UPP 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
This command defines the duration of the level measurement used to determine the
optimal reference level automatically (for
SENS:ADJ:LEV ON
).
Parameters:
<Duration>
<numeric value> in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:LEV:DUR:5
Manual operation:
See
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
<Mode>
This command selects the way the R&S
FSVA/FSV determines the length of the mea-
surement that is performed while determining the ideal reference level.
Parameters:
<Mode>
AUTO
Automatically determines the measurement length.
MANual
Manual definition of the measurement length.
*RST:
AUTO
Example:
ADJ:CONF:LEV:DUR:MODE MAN
Specifies manual definition of the measurement duration.
ADJ:CONF:LEV:DUR:5
Specifies the duration manually.
[SENSe:]ADJust:LEVel
This command initiates automatic setting of the RF attenuation to the level of the
applied signal.
Note:
The following command must be synchronized to the end of the autorange proc-
ess using *WAI, *OPC oder *OPC?, because otherwise the autorange process will be
stopped.
Example:
ADJ:LEV
Adjusts the reference level to the current measurement.
Usage:
Event
Mode:
VSA
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...