Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
280
Operating Manual 1176.7578.02 ─ 09
4.10 OUTPut Subsystem
OUTPut:DIQ
<State>
If enabled, the captured IQ data is output to the R&S Digital I/Q Interface in a continu-
ous stream. This function requires the LVDS interface option (R&S
FSV-B17).
Digital input and digital output cannot be used simultaneously.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
OUTP:DIQ ON
Mode:
ADEMOD, IQ, VSA
OUTPut:DIQ:CDEVice
This command queries the current configuration and the status of the digital baseband
output to the optional R&S Digital I/Q Interface (option R&S
FSV-B17).
For details see the R&S Digital I/Q Interface description for the base unit.
Return values:
<ConnState>
Defines whether a device is connected or not.
0
No device is connected.
1
A device is connected.
<DeviceName>
Device ID of the connected device
<SerialNumber>
Serial number of the connected device
<PortName>
Port name used by the connected device
<NotUsed>
to be ignored
<MaxTransferRate> Maximum data transfer rate of the connected device in Hz
<ConnProtState>
State of the connection protocol which is used to identify the
connected device.
Not Started
Has to be Started
Started
Passed
Failed
Done
OUTPut Subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...