Note
: switch to single sweep mode (
INIT:CONT OFF
) before importing I/Q data as
otherwise the instrument will continue to measure data and display the current results
rather than the imported data.
Parameters:
<FileName>
Complete file name including the path
Example:
MMEM:LOAD:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Loads I/Q data from the specified file.
Usage:
Setting only
MMEMory:SELect:ITEM:VIQData
<Mode>
If enabled, the captured I/Q data is included in the save set when instrument data is
stored (single sweep mode only).
Parameters:
<Mode>
ON | OFF
Mode:
VSA
MMEMory:STORe:IQ:STATe
1, <FileName>
This command stores the complex I/Q data to the specified
.iq.tar
file in 32-bit float-
ing point format.
Parameters:
<FileName>
Complete file name including the path
Example:
MMEM:STOR:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Stores I/Q data to the specified file.
Manual operation:
See
MMEMory:STORe:IQ:COMM
<Description>
Defines a description of the export file which is stored with the data and also displayed
in the file selection dialog box for I/Q data import and export.
Parameters:
<Description>
Example:
MMEM:STOR:IQ:COMM 'Device test 1b'
Creates a description for the export file.
MMEM:STOR:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Stores I/Q data and the comment to the specified file.
Manual operation:
See
MMEMory Subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...