Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
255
Operating Manual 1176.7578.02 ─ 09
Setting parameters:
<FitOffset>
numeric value
Range:
-8000 to 8000
*RST:
0
Default unit: SYM
Example:
CALC:TRAC:ADJ:ALIG:OFFS 5
The display range is shifted by 5 symbols towards the right.
Mode:
VSA
Manual operation:
See
CALCulate<n>:TRACe<t>:ADJust:ALIGnment[:DEFault]
<Alignment>
This command defines where the relevant event (reference point) is to appear in the
result range.
Suffix:
<n>
.
1..4
irrelevant
<t>
1..6
irrelevant
Setting parameters:
<Alignment>
LEFT | CENTer | RIGHt
LEFT
The reference point is displayed at the left edge of the result
range.
CENTer
The reference point is displayed in the middle of the result
range.
RIGHt
The reference point is displayed at the right edge of the result
range.
*RST:
LEFT
Example:
CALC:TRAC:ADJ:ALIG LEFT
The reference point is displayed at the left edge.
Mode:
VSA
Manual operation:
See
CALCulate<n>:TRACe<t>:ADJust[:VALue]
<Reference>
This command defines the reference point for the display.
Suffix:
<n>
.
1..4
irrelevant
CALCulate subsystem