Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
251
Operating Manual 1176.7578.02 ─ 09
CALCulate<n>:STATistics:PRESet
This command sets both axis of the statistics measurement to measurement depend-
ent default values.
Suffix:
<n>
.
1..4
Example:
CALC:STAT:PRES
Usage:
Event
Mode:
VSA
Manual operation:
See
CALCulate<n>:STATistics:SCALe:AUTO
<AutoMode>
Sets the x-axis of the statistics measurement depending on the measured values.
Suffix:
<n>
.
1..4
Setting parameters:
<AutoMode>
ONCE
Example:
CALC3:STAT:SCAL:AUTO ONCE
Usage:
Setting only
Mode:
VSA
Manual operation:
See
CALCulate<n>:STATistics:SCALe:X:BCOunt
<StatisticsNofColumns>
This command defines the number of columns for the statistical distribution.
Suffix:
<n>
.
1..4
Setting parameters:
<StatisticsNofColumns>
numeric value
Range:
2 to 1024
*RST:
101
Default unit: NONE
Example:
CALC:STAT:SCAL:X:BCO 10
Sets the number of columns to 10.
Mode:
VSA
Manual operation:
See
CALCulate subsystem