Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:MERR:RPE:VAL 2.4%
define a limit of 2.4%
Mode:
VSA
CALCulate<n>:LIMit:MACCuracy:OOFFset:CURRent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:OOFFset:MEAN:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:OOFFset:PEAK:VALue
<LimitValue>
This command defines the upper limit for the current, peak or mean I/Q offset. Note
that the limits for the current and the peak value are always kept identical.
Suffix:
<n>
.
1..4
window
Setting parameters:
<LimitValue>
numeric value
Range:
-200.0 to 0.0
*RST:
-40.0 (mean: -45.0)
Default unit: dB
Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:OOFF:PEAK:VAL -50dB
define a limit of -50 dB
Mode:
VSA
CALCulate<n>:LIMit:MACCuracy:PERRor:PCURrent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:PERRor:PMEan:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:PERRor:PPEak:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:PERRor:RCURrent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:PERRor:RMEan:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:PERRor:RPEak:VALue
<LimitValue>
This command defines the value for the current, peak or mean phase error (peak or
RMS) limit. Note that the limits for the current and the peak value are always kept iden-
tical.
Suffix:
<n>
.
1..4
window
Setting parameters:
<LimitValue>
numeric value
the value x (x>0) defines the interval [-x; x]
Range:
0.0 to 360
*RST:
3.5 (RMS: 1.5)
Default unit: deg
CALCulate subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...