Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
194
Operating Manual 1176.7578.02 ─ 09
4. Press "Remove from Standard".
The pattern is deleted and removed from the list of "Standard Patterns", but is still
available for assignment from the list of "All Patterns".
3.3.7 Working With Known Data Files
For various vector signal analysis functions the measured signal is compared to a
defined ideal reference signal. The more precise the reference signal, the more precise
the results become. In the best case, the possible data sequences within the signal to
analyze are known in advance and can be used to compare the measured data to.
This is similar to defining a pattern for the entire result range. Thus, a falsely estimated
reference signal (due to false symbol decisions) is avoided and does not influence the
error calculation.
As of firmware version R&S
FSVA/FSV 1.70, you can load xml files containing the pos-
sible sequences to the R&S
FSV-K70 application and use them to compare the mea-
sured data to. In particular, you can use known data for the following functions:
●
Fine synchronization during the demodulation process (see
and
●
Calculation of the Bit Error Rate (BER), see
Chapter 3.1.1.23, "Bit Error Rate
3.3.7.1
Dependencies and Restrictions when Using Known Data
When you use Known Data files as a reference, some dependencies to other settings
and restrictions for other functions apply.
Modulation Order
The "Modulation Order" selected in the "Modulation" settings in the R&S
FSV-K70
application must correspond to the modulation order value specified in the xml file
(<ModulationOrder> element).
Demodulation
Demodulation using synchronization to the Known Data may increase the measure-
ment duration, as each detected symbol must be compared to each possible sequence
in the data file.
Result Length
The "Result Length" specified in the "Result Range" dialog box in the R&S
FSV-K70
application must be identical to the length of the specified symbol sequences in the xml
file (<ResultLength> element).
Result Range Alignment
●
Bursted signals
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...