Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
184
Operating Manual 1176.7578.02 ─ 09
CALC:TRAC:ADJ:ALIG LEFT
//Aligns the result range to the left edge of the capture buffer
CALC:TRAC:ADJ:ALIG:OFFS 1
//Defines an offset of 1 symbol from the capture buffer start
DISP:TRAC:X:VOFF 1
//Defines the symbol number 1 as the capture buffer start
3.3.3 Changing the Display Scaling
Depending on the type of display (time, spectrum or statistics), various scaling func-
tions are available to adapt the result display to the current data. Scaling functions are
located in the "Ranges" submenu of the "Amplitude" menu (see
keys of the Amplitude Menu (R&S
3.3.3.1
Scaling Time and Spectrum Diagrams
The range of the displayed y-axis for time and spectral diagrams can be defined in the
following ways:
●
manually, by defining the range size, reference values and positions
●
automatically, according to the current results
To define the scaling manually using a reference point
With this method, you define a reference value and a position at which this value is to
be displayed on the y-axis.
1. Focus the result screen.
2. Select "AMPT > Ranges > Y-Axis Reference Value" (see
3. Enter a reference value for the y-axis in the current unit.
4. Select "AMPT > Ranges > Y-Axis Reference Position" (see
5. Enter the position at which this value is to be displayed on the y-axis. The position
is a percentage of the entire length, where 100 % refers to the top edge.
6. Select "AMPT > Ranges > Y-Axis Range" (see
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...