Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
18
Operating Manual 1176.7561.02 ─ 06
Max Hold
The maximum value is determined over several sweeps and displayed. The
R&S
FSVA/FSV saves the sweep result in the trace memory only if the new value is
greater than the previous one.
The detector is automatically set to "Positive Peak".
This mode is especially useful with modulated or pulsed signals. The signal spectrum
is filled up upon each sweep until all signal components are detected in a kind of enve-
lope.
This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE MAXH
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Min Hold
The minimum value is determined from several measurements and displayed. The
R&S
FSVA/FSV saves the smallest of the previously stored/currently measured values
in the trace memory.
The detector is automatically set to "Negative Peak".
This mode is useful e.g. for making an unmodulated carrier in a composite signal visi-
ble. Noise, interference signals or modulated signals are suppressed whereas a CW
signal is recognized by its constant level.
This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE MINH
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Average
The average is formed over several sweeps. The
determines the number
of averaging procedures.
All available detectors can be selected. If the detector is automatically selected, the
sample detector is used (see
Chapter 2.1.8, "Detector Overview"
This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE AVER
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
View
The current contents of the trace memory are frozen and displayed.
Note:
If a trace is frozen, the instrument settings, apart from level range and reference
level (see below), can be changed without impact on the displayed trace. The fact that
the displayed trace no longer matches the current instrument setting is indicated by the
icon on the tab label.
If the level range or reference level is changed, the R&S
FSVA/FSV automatically
adapts the measured data to the changed display range. This allows an amplitude
zoom to be made after the measurement in order to show details of the trace.
Instrument Functions Analog Demodulation