Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
166
Operating Manual 1176.7561.02 ─ 06
Example:
TRIG:LEV:IFP -30DBM
Manual operation:
See
TRIGger<n>[:SEQuence]:LEVel:RFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event. Note that any RF attenuation or preamplification is
considered when the trigger level is analyzed. If defined, a reference level offset is also
considered.
Suffix:
<n>
.
irrelevant
Parameters:
<TriggerLevel>
*RST:
-20 dBm
Example:
TRIG:LEV:RFP -30dBm
TRIGger<n>[:SEQuence]:LEVel:PM
<Level>
The command sets the level when PM-modulated signals are used as trigger source.
For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful,
the measurement time must cover at least 5 periods of the audio signal.
Suffix:
n
.
irrelevant
Parameters:
<Level>
Range:
-1000 to +1000
*RST:
0 RAD
Default unit: RAD | DEG
Example:
TRIG:LEV:PM 1.2 RAD
Sets the PM trigger threshold to 1.2 rad
Mode:
ADEMOD
TRIGger<n>[:SEQuence]:TIME:RINTerval
<Interval>
This command sets the repetition interval for the time trigger source.
Suffix:
<n>
.
irrelevant
Parameters:
<Interval>
2.0 ms to 5000
*RST:
1.0
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Remote Commands of the Analog Demodulation (R&S