DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
.........................................................112
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................... 112
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
........................................................113
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
............................................113
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
....................................................113
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue
.......................................................114
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
...................................................................114
DISPlay:MTABle
<DisplayMode>
This command turns the marker table on and off.
Parameters:
<DisplayMode>
ON
Marker table is displayed.
OFF
Marker table is not displayed.
AUTO
Marker table is only displayed if 2 or more markers are active.
*RST:
AUTO
Example:
To activate the table display:
DISP:MTAB ON
To query the current state of the marker table display:
DISP:MTAB?
Manual operation:
See
DISPlay[:WINDow<n>]:STATe
<State>
This command activates the measurement specified window.
Suffix:
<n>
.
1...4
window
Parameters:
<State>
ON | OFF
Example:
DISP:WIND2:STAT ON
Displays a second window (Screen B).
Usage:
SCPI confirmed
Mode:
ADEMOD
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<1|2>]:SELect
Moves the focus area to the selected window and subwindow.
Remote Commands of the Analog Demodulation (R&S