Remote Commands
R&S
®
FSMR3000
587
User Manual 1179.0116.02 ─ 02
........................................................................................588
............................................................................... 588
.................................................................................... 588
FORMat:DEXPort:HEADer
<State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
"Include Instrument & Measurement Settings"
MMEMory:STORe<n>:LIST
<FileName>
This command exports the SEM and spurious emission list evaluation to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:LIST 'test'
Stores the current list evaluation results in the
test.dat
file.
MMEMory:STORe<n>:PEAK
<FileName>
This command exports the marker peak list to a file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Suffix:
<n>
.
Managing Settings and Results