Remote Commands
R&S
®
FSMR3000
427
User Manual 1179.0116.02 ─ 02
OUTPut<up>:ADEMod[:ONLine][:STATe]
<State>
This command enables or disables online demodulation output to the IF/VIDEO/
DEMOD output connector on the rear panel of the R&S
FSMR3.
Suffix:
<up>
.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
OUTP:ADEM ON
Manual operation:
See
"Online Demodulation Output State"
OUTPut<up>:ADEMod[:ONLine]:SOURce
<WindowName>
This command selects the result display whose results are output. Only active time
domain results can be selected.
Suffix:
<up>
.
Parameters:
<WindowName>
<string>
String containing the name of the window.
By default, the name of a window is the same as its index. To
determine the name and index of all active windows, use the
FOCus
Dynamically switches to the currently selected window. If a win-
dow is selected that does not contain a time-domain result dis-
play, the selection is ignored and the previous setting is main-
tained.
Example:
OUTP:ADEM:ONL:SOUR 'AnalogDemod'
OR:
DISP:WIND1:SEL
OUTP:ADEM:SOUR FOC
Manual operation:
See
OUTPut<up>:ADEMod[:ONLine]:AF[:CFRequency]
<Frequency>
This command defines the cutoff frequency for the AC highpass filter (for AC coupling
only, see
Suffix:
<up>
.
Configuring the Measurement