Remote Commands
R&S
®
FSMR3000
395
User Manual 1179.0116.02 ─ 02
Manual operation:
See
See
Table 12-2: Available channel types and default channel names
Application
<ChannelType> Parameter
Default Channel Name*)
Measuring Receiver
MRECeiver
Measuring Receiver
Spectrum (R&S
FSMR3-
B1)
SANalyzer
Spectrum
I/Q Analyzer (R&S
FSMR3-
B1)
IQ
IQ Analyzer
Phase Noise
(R&S
FSMR3-B60)
PNOise
Phase Noise
Pulse (R&S
FSMR3-K6)
PULSE
Pulse
Avionics (R&S
FSMR3-
K15)
AVIonics
Avionics
Vector Signal Analysis
(VSA, R&S
FSMR3-K70)
DDEM
VSA
Note: the default channel name is also listed in the table. If the specified name for a new channel already
exists, the default name, extended by a sequential number, is used for the new channel.
INSTrument:REName
<ChannelName1>, <ChannelName2>
This command renames a channel.
Setting parameters:
<ChannelName1>
String containing the name of the channel you want to rename.
<ChannelName2>
String containing the new channel name.
Note that you cannot assign an existing channel name to a new
channel; this will cause an error.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Example:
INST:REN 'Measuring Receiver 2','Measuring
Receiver 3'
Renames the channel with the name 'Measuring Receiver 2' to
'Measuring Receiver 3'.
Usage:
Setting only
Manual operation:
See
INSTrument[:SELect]
<ChannelType> | <ChannelName>
This command activates a new channel with the defined channel type, or selects an
existing channel with the specified name.
Also see
Selecting the Operating Mode and Application