Common Analysis and Display Functions
R&S
®
FSMR3000
223
User Manual 1179.0116.02 ─ 02
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Blank
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Trace 2:
Average
Trace 3:
Min Hold
Blank
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Trace 2:
Clear Write
Trace 3:
Min Hold
Blank
8.4.1.3
How to Configure a Standard Trace
Step-by-step instructions on configuring the trace settings are provided here.
Trace settings are configured in the "Traces" dialog box.
To display the "Traces" dialog box, do one of the following:
●
Select "Analysis" from the "Overview", then select the "Traces" tab.
1. For each trace, select the "Trace Mode" and "Trace Detector". Traces with the
trace mode "Blank" are not displayed.
2. To configure several traces to predefined display modes in one step, press the but-
ton for the required function:
● "Preset All Traces"
● "Set Trace Mode Max | Avg | Min"
● "Set Trace Mode Max | ClrWrite | Min"
3. For "Average" trace mode, define the number of measurements to be averaged in
the "Count:" field.
4. If linear scaling is used, select the "Average Mode": "Linear".
5. To improve the trace stability, increase the number of "Sweep Points" or the
"Sweep Time" (in the "Sweep" settings).
All configured traces (not set to "Blank") are displayed after the next measurement.
Trace Configuration