Common Measurement Settings
R&S
®
FSMR3000
125
User Manual 1179.0116.02 ─ 02
Generator
type
Model
Driver file
TTL sup-
port
Generator type
Model
Driver file
TTL sup-
port
SMGL
all
SMGL
-
2
GHz
SMY02
-
SMGU
all
SMGU
-
SMH
all
SMH
-
SMHU
SMHU
-
SMIQ
2
GHz
SMIQ02
X
2
GHz
SMIQ02B
X
2
GHz
SMIQ02E
-
3
GHz
SMIQ03
X
3
GHz
SMIQ03B
X
3
GHz
SMIQ03E
-
4
GHz
SMIQ04B
X
6
GHz
SMIQ06B
X
1) Requires firmware version V2.10.x or later on the signal generator
2) Requires firmware version V1.10.x or later on the signal generator
3) Requires the option SMR-B11 on the signal generator
4) Requires firmware version V3.20.200 or later on the signal generator
Generator Setup Files
For each signal generator type to be controlled by the R&S
FSMR3 a generator setup
file must be configured and stored on the R&S
FSMR3. The setup file defines the fre-
quency and power ranges supported by the generator, as well as information required
for communication. For the signal generators listed in
"Overview of Supported Genera-
on page 123, default setup files are provided. If necessary, these files can be edi-
ted or duplicated for varying measurement setups or other instruments.
The existing setup files can be displayed in an editor in read-only mode directly from
the "External Generator" configuration dialog box. From there, they can be edited and
stored under a different name, and are then available on the R&S
FSMR3.
Coupling the Frequencies
●
Manual coupling
: a single frequency is defined
●
Automatic coupling
: a series of frequencies is defined (one for each sweep
point), based on the current frequency at the RF input of the R&S
FSMR3; the RF
frequency range covers the currently defined span of the R&S
FSMR3 (unless limi-
ted by the range of the signal generator)
Automatic coupling
If automatic coupling is used, the output frequency of the generator (source frequency)
is calculated as follows:
Input and Frontend