Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
94
User Manual 1179.0880.02 ─ 03
Suffix:
<n>
.
<w>
irrelevant
<t>
irrelevant
Parameters:
<State>
OFF | 0
Selects manual scaling of the diagram.
ON | 1
Automatically scales the diagram when new results are availa-
ble.
ONCE
Automatically scales the diagram once whenever required.
*RST:
ON
Example:
//Scale the axis each time new results are available
DISP:TRAC:X:AUTO ON
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:AUTO
<ONCE>
This command automatically scales the y-axis of a diagram based on the displayed
results.
Suffix:
<n>
.
<w>
<t>
irrelevant
Setting parameters:
<ONCE>
ALL
Scales the y-axis in all windows for an ideal viewing experience.
DEFault
Restores the default scale of the y-axis.
ONCE
Scales the y-axis in a specific window for an ideal viewing expe-
rience.
Example:
//Automatically scale the y-axis in subwindow 2 of window 2
DISP:WIND2:SUBW2:TRAC:Y:AUTO ONCE
Usage:
Setting only
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Z[:SCALe]:AUTO
<State>
This command turns automatic scaling of the x-axis in graphical result displays on and
off.
Analysis