Quick Start Guide 1302.6163.62 - 04
2.17
R&S ESU
Preparing for Use
Connecting External Devices
➢
Press the
ENTER
key.
The list of available generators will open.
➢
Using the rotary knob, select the desired generator and activate
ENTER
by pressing the
rotary knob.
The generator list will close, and the selected generator will then appear in the
SELECT
GENERATOR
table. Simultaneously, the limits for frequency and output power which
you can set on the generator will appear in the F MIN, F MAX, P MIN and P MAX fields.
➢
Using the
U
r
key, move the selection bar to the IFC column and press the
ENTER
key.
The list of available control interfaces will appear.
➢
If the TTL interfaces from the R&S ESU and generator are connected, select
TTL
by
using the rotary knob and then activate it by pressing the rotary knob.
➢
If no TTL interface is provided or if the TTL interfaces are not connected, select
GPIB
by
using the rotary knob and then activate it by pressing the rotary knob.
➢
Using the
U
r
key, move the selection bar to the GPIB ADDR column and press the
ENTER key.
The entry field for the GPIB address of the generator will open.
➢
Enter the GPIB address of the generator and confirm with
ENTER
.
The specified address will be added to the table.
3. Select the reference frequency for the generator:
➢
If you want to operate the generator using its internal reference, change the
GEN REF
softkey to the state
INT
.
➢
If you want to operate the generator using an external reference, change the
GEN REF
softkey to the state
EXT
.
This concludes the procedure for defining the default setting for the generator.
2.3.6.3 Activating the Generator Configuration and Setting the Output Level
To select the active generator configuration, press the
FREQUENCY SWEEP
softkey. The
FREQUENCY SWEEP
table will open, and the selection bar will be located under the
STATE
column in the
SRC 1
row.
➢
If you want to use the
SRC 1
generator configuration, insert a checkmark in the field by
pressing the
ENTER
key.
A second generator configuration can be inserted in advance by
repeating the settings in the line SRC 2. This makes it easy to switch
from one generator type to another.
Содержание esu
Страница 1: ...R S ESU EMI Test Receiver Quick Start Guide Quick Start Guide Test Measurement 1302 6163 62 04 ...
Страница 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Страница 28: ......
Страница 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Страница 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Страница 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Страница 175: ...Quick Start Guide 1302 6163 62 04 6 33 R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop ...
Страница 178: ...R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop 6 36 Quick Start Guide 1302 6163 62 04 ...
Страница 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Страница 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...