R&S ESU
Front and Rear Panel
Rear Panel Connections
1.16
Quick Start Guide 1302.6163.62 - 04
20.4 MHz IF Output (IF OUT 20.4 MHz)
The 20.4 MHz IF signal of the R&S ESU is provided at the 20.4 MHz OUT female BNC
connector. For resolution bandwidths between 100 kHz and 10 MHz, the bandwidth
corresponds to the selected bandwidth. For bandwidths
≤
100 kHz, the bandwidth of the output
is equal to 2.6 * resolution bandwidth, where the minimum value is 2.6 kHz (non-FFT). In the
analyzer mode, the level at the IF output in the case of a signal at the reference level is 0 dBm
if the resolution bandwidth is
≥
100 kHz; if the resolution bandwidth is <100 kHz, the level is
-10 dBm (for mixer levels
≥
-60 dBm).
Video Output (VIDEO OUT)
In analyzer mode, the voltage of the analog detector (video) is output via the BNC connector
VIDEO OUT.
The video signal is unfiltered, i.e. the video bandwidth (VBW) setting on the instrument has no
effect on the video output. The usable video bandwidth is limited only by the set resolution
bandwidth (RBW) to approximately ½ * RBW.
The output voltage is always logarithmic, independent of the selected scaling on the display.
Likewise the selected detectors (e.g. RMS) affect only the display but not the video output. The
output impedance is 50
Ω
. All subsequent voltage values are typical values and apply to the
output without load.
With RBW = 100 kHz as well as with all root raised cosine (RRC) and channel filters no video
signal is available.
RBW 200 kHz to 10 MHz
Output voltage at signal level = reference level: 1.8 V
Slope14 mV / dB
Exception: Ref_Level / dBm - RF_Attenuation / dB < -50
Output voltage at signal level = reference level:
1.8 V + (Ref_Level / dBm - RF_Attenuation / dB + 50) * 14 mV
TG I / AM IN; TG Q / FM IN (Option R&S FSU-B9)
The two female connectors TG I /AM IN and TG Q /FM IN are used to modulate the tracking
generator (option R&S FSU-B9) by means of an external signal.
The input voltage range is ±0.5 V; the input impedance is 50
Ω
.
Use suitable double shielded cables.
Содержание esu
Страница 1: ...R S ESU EMI Test Receiver Quick Start Guide Quick Start Guide Test Measurement 1302 6163 62 04 ...
Страница 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Страница 28: ......
Страница 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Страница 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Страница 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Страница 175: ...Quick Start Guide 1302 6163 62 04 6 33 R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop ...
Страница 178: ...R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop 6 36 Quick Start Guide 1302 6163 62 04 ...
Страница 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Страница 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...