Remote Control – Description of Commands
INPut Subsystem
6.132
Operating Manual 1302.6163.12 - 03
INPut<1|2>:IMPedance
50 | 75
This command sets the nominal input impedance of the instrument. The set impedance is taken into
account in all level indications of results.
The setting 75
Ω
should be selected, if the 50
Ω
input impedance is transformed to a higher impedance
using a 75
Ω
adapter of the RAZ type (= 25
Ω
in series to the input impedance of the instrument). The
correction value in this case is 1.76 dB = 10 log (75
Ω
/ 50
Ω
).
Example:
"INP:IMP 75"
Characteristics:
*RST value: 50
Ω
SCPI: conform
Mode:
R, A
INPut<1|2>:MIXer[:POWer]
<numeric value>
This command defines the desired power at the input mixer of the R&S ESU. On any change to the
reference level the RF attenuation will be adjusted in a way that makes the difference between
reference level and RF attenuation come as close to the desired mixer level as possible.
Example:
"INP:MIX -30"
Characteristics:
*RST value: - 25 dBm
SCPI: device-specific
Mode:
A
INPut<1|2>:MIXer:AUTO
ON | OFF
This command enables/disables the automatic setup of the mixer level.
Example:
"INP:MIX:AUTO ON"
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
INPut<1|2>:PRESelection[:STATe]
ON | OFF
In analyzer mode, this command switches the preselection on or off.
Example:
":INP:PRES:STAT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
In receiver mode, the preselection is always switched on and cannot be switched
off.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...