Operating Manual 1302.6163.12 - 03
4.221
R&S ESU
Instrument Functions
Analyzer Mode
MARGIN
The
MARGIN
softkey activates entry of the margin, i.e. the acceptance threshold for
determining the peak list. The limit line is shifted by this value when the maximums
are determined. The value range extends from -200 dB to 200 dB. The default value
is 6 dB.
Remote command:
CALC:PEAK:MARG –200dB...200dB
VIEW PEAK
LIST
The
VIEW PEAK LIST
softkey opens the submenu for viewing the peak list.
It is activated for display only after a PEAK search has been performed with
PEAK
SEARCH.
If no limit check is active, a delta limit of +200dB is listed.
Remote command:
TRACe? SPURious
SORT BY
FREQUENCY
The
SORT BY FREQUENCY
softkey sorts the table according to the entries in the
FREQUENCY column in descending order.
Remote command:
--
SORT BY
DELTA LIM
The
SORT BY DELTA LIM
softkey sorts the table according to the entries in the
DELTA LIM column (default) in descending order. If no limit line has been specified,
an offset of 200 dB is assumed for all peaks.
Remote command:
--
ASCII FILE
EXPORT
The
ASCII FILE EXPORT
softkey stores the peak list in ASCII format to a file on the
flash disk or a memory stick.
The file has a header containing important parameters for scaling, several data sec-
tions containing the sweep settings per range and a data section containing the
peak list.
The header data is made up of three columns, separated by ';':
parameter name; numeric value; base unit
VIEW PEAK LIST
TRACE / Detector
FREQUENCY
LEVEL dBm
DELTA LIMIT dB
1 RMS
80.0000 MHz
-36.02
-5.02
1 RMS
80.0001 MHz
-30.07
+0.24
1 RMS
85.1234 MHz
-30.02
-0.02
1 AVERAGE
130.234 MHz
-29.12
-5.12
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...