Preparing For Use
R&S
®
ESR
42
Getting Started 1316.3749.02 ─ 09
5.15
Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. In
order to check whether the installed options correspond to the options indicated
on the delivery note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Info" softkey.
3. Press the "Ve Options" softkey.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.
For an overview of the available options, refer to the datasheet.
5.16
Performing a Self Alignment and a Self Test
Operating temperature
Before performing this functional test, make sure that the instrument has
reached its operating temperature (for details, refer to the data sheet).
Performing a self alignment
1. Press the [SETUP] key.
2. Press the "Alignment" softkey.
3. Press the "Self Alignment" softkey.
Once the system correction values have been calculated successfully, a mes-
sage is displayed.
To display the alignment results again later
●
Press the [SETUP] key.
●
Press the "Alignment" softkey.
●
Press the "Show Align Results" softkey.
Performing a Self Alignment and a Self Test