Remote Control
R&S
®
ESR
686
User Manual 1175.7068.02 ─ 12
The command is available for the R&S
ENV216 network.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:LISN:FILT:HPAS ON
Turns on the highpass filter.
Manual operation:
See
See
INPut:LISN:PHASe
<Phase>, [<Phase>]
This command selects the phase of the network used for the premeasurement.
Parameters:
<Phase>
L1
L2
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
L3
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
N
*RST:
L1
Example:
INP:LISN:PHAS L1,L2
Selects phases L1 and L2 for the scan.
Manual operation:
See
See
INPut:LISN[:TYPE]
<Type>
This command turns automatic control of the network used for the premeasurement on
and off. It also selects the type of network in use.
Parameters:
<Type>
ENV216
R&S ENV 216: two phases and highpass are controllable.
ENV432
R&S ENV 432: four phases are controllable.
ENV4200
R&S ENV 4200: four phases are controllable.
ESH2Z5
R&S ESH2-Z5: four phases and protective earth are controlla-
ble.
ESH3Z5
R&S ESH3-Z5: two phases and protective earth are controllable.
Remote Commands in Receiver Mode