Spectrum Measurements
R&S
®
ESR
406
User Manual 1175.7068.02 ─ 12
Remote command:
Color Mapping ← Spectrogram
Displays the "Color Mapping" dialog box to configure the display of the spectrogram
(assignment of colors to power levels).
For more information see
Remote command:
DISPlay:WINDow:SGRam:COLor:LOWer
DISPlay:WINDow:SGRam:COLor:UPPer
Time Stamp (On Off) ← Spectrogram
Activates and deactivates the time stamp. The time stamp shows the system time
while the measurement is running. In single sweep mode or if the sweep is stopped,
the time stamp shows the time and date of the end of the sweep.
When active, the time stamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SGRam:TSTamp[:STATe]
Clear Spectrogram ← Spectrogram
Resets the Spectrogram result display and clears the history buffer.
Remote command:
CALCulate<n>:SGRam:CLEar[:IMMediate]
6.3.3
Markers
6.3.3.1
Controlling Markers
The markers are used for marking points on traces, reading out measurement results
and for selecting a display section quickly. The R&S
ESR provides 16 markers per
trace.
To open the Marker menu
●
Press the MKR key.
The "Marker" menu is displayed. If no marker is active, marker 1 is activated and a
peak search on the trace is carried out. Otherwise, the edit dialog box for the last
activated marker is opened and the current frequency/time value is displayed.
Further information
●
"Displayed Marker Information"
●
Chapter 5.4.2.2, "Positioning Markers"
Analysis