R&S ENY-FTS
Operation
Operating Manual 1309.8949.14 - 0
2
7
2 Operation
2.1 Functional test in accordance with CISPR 16-1-2
The R&S ENY-FTS in connection with a network analyzer allows the functional testing
of the impedance stabilization networks (ISN) R&S ENY21, R&S ENY41, R&S ENY81
and R&S ENY81-CA6. This functional testing includes the verification of the
asymmetrical impedance and phase, voltage division factor, longitudinal conversion
loss and decoupling attenuation.
Requirements and examples for the measurement of the parameters of an ISN are
defined in Annex E of the basic standard CISPR 16-1-2. The ITU-T recommendations
G.117 and O.9 are offering further background information for measurements on
symmetrical telecommunication lines. For this purpose the R&S ENY-FTS contains a
LCL balance measuring bridge with a balanced output of 100
C
(1 mm banana socket)
and various adapters for the reference calibration of the bridge and the connection of
the measuring bridge to the ISN.
Fig. 1
Parts of the R&S ENY-FTS
Jig CAL
U100B
Jig LCL-Probe
Jig TRA U0-150
SAR T801
SAR T2x0
2x SAR T802
SAR T4x0
SAR T602
SAR T8x0
SAR M116
Jig IMA U0
A-800
A-150
A-OPEN
A-SHORT
A-801
2x Cable
A-8-0-P1 to P4
A-8-CAT6-P1 to P4
A-8-CAT6-P1x to P4x
A-2-0-RJ45
A-2-0-RJ11
A-TERMINATION
A-CALLOAD
BMB-1000
A-REFLOAD
A-50
Содержание ENY21
Страница 14: ...R S ENY FTS Inhaltsverzeichnis Betriebshandbuch 1309 8949 14 02 4...
Страница 50: ...R S ENY FTS Contents Operating Manual 1309 8949 14 02 4...
Страница 66: ......