Maintenance
R&S
®
DVMS1
51
Getting Started 2113.7619.02 ─ 12
9
Maintenance
The R&S
DVMS does not need periodic maintenance. Clean the outside of the instru-
ment when necessary and check the rated data from time to time.
If any problem arises, contact one of our customer support centers. The addresses of
our customer support centers are provided at the beginning of this manual.
9.1
Cleaning the Instrument
Shock hazard
Before cleaning the R&S
DVMS, make sure that the R&S
DVMS is switched off and
disconnected from all power supplies.
1. Clean the outside of the R&S
DVMS using a soft, lint–free dust cloth.
2. Make sure that vents are unobstructed.
Instrument damage caused by cleaning agents
Cleaning agents contain substances that can damage the R&S
DVMS, e.g. cleaning
agents that contain a solvent can damage the front panel labeling or plastic parts.
Never use cleaning agents such as solvents (thinners, acetone, etc.), acids, bases, or
other substances.
9.2
Replacing the Fuses
The R&S
DVMS is protected by two fuses (IEC
127
– T
2.0
H/250
V) on the rear panel
next to the AC power switch.
Shock hazard
Before replacing a fuse, make sure that the R&S
DVMS is switched off and disconnec-
ted from all power supplies.
Always use fuses supplied by Rohde & Schwarz as spare parts, or fuses of the same
type and rating.
Replacing the Fuses