LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
783
User Manual 1173.9628.02 ─ 16
If you get firmware errors even with the properly installed latest software version,
please send a problem report including log files to Rohde & Schwarz.
●
102 (Unidentified Error)
:
Indicates an error not covered by other reliability values. For troubleshooting
please follow the steps described for "101 (Firmware Error)".
●
103 (Parameter Error)
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com
4.5.2 General Measurement Settings
The commands valid for all LTE measurements are described here:
"General Measurement Settings"
4.5.3 PRACH Measurement Commands
The commands for the "LTE PRACH" measurement are divided into the groups listed
below.
Measurement Control and States
......................................................................... 784
..................................................................................786
Measurement Parameters - General Part
.............................................................788
Modulation Measurement Settings
........................................................................791
...............................................................................794
................................................................................................797
.............................................................................. 799
EVM vs. Subcarrier Results (Traces)
....................................................................802
Magnitude Error Results (Traces)
.........................................................................803
................................................................................803
EVM vs. Preamble Results (Traces)
.....................................................................804
Power vs. Preamble Results (Traces)
...................................................................804
I/Q Constellation Results (Traces)
........................................................................ 805
Power Dynamics Results (Traces)
........................................................................806
"TX Measurement" Results (Single Values)
..........................................................806
EVM / Power vs. Preamble Results (Single Values)
.............................................809
Power Dynamics Results (Single Values)
.............................................................811
Command Reference
135-1095-0799