![R&S CMW-KM5 Series Скачать руководство пользователя страница 693](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802693.webp)
LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
693
User Manual 1173.9628.02 ─ 16
Example:
See
Usage:
Query only
Firmware/Software:
V2.0.10
V2.0.20: CALCulate commands
V2.1.25: increased maximum number of segments to 250
V2.1.30: increased maximum number of segments to 512
V3.0.50: increased maximum number of segments to 1000
Options:
R&S CMW-KM012
FETCh:LTE:MEAS<i>:MEValuation:LIST:SEGMent<no>:ESFLatness:CURRent:
SCINdex?
Return subcarrier indices of the equalizer spectrum flatness measurement for segment
<no> in list mode. At these SC indices the current minimum and maximum power of
the equalizer coefficients have been detected within range 1 and range 2.
Suffix:
<no>
.
1..1000
Return values:
<Reliability>
In list mode, a zero reliability indicator indicates that the results
in all measured segments are valid. A non-zero value indicates
that an error occurred in at least one of the measured segments.
<SegReliability>
Reliability indicator for the segment. The meaning of the
returned values is the same as for the common reliability indica-
tor, see previous parameter.
<StatistExpired>
Reached statistical length in slots
Range:
0 to 1000
<OutOfTolerance>
Percentage of measured subframes with failed limit check
Range:
0 % to 100 %
Default unit: %
<Maximum1>
SC index of Max (Range 1)
Range:
0 to 1199
<Minimum1>
SC index of Min (Range 1)
Range:
0 to 1199
<Maximum2>
SC index of Max (Range 2)
Range:
0 to 1199
<Minimum2>
SC index of Min (Range 2)
Range:
0 to 1199
Usage:
Query only
Command Reference
深圳德标仪器