LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
626
User Manual 1173.9628.02 ─ 16
Parameters:
<ACLRstatistics>
Statistical length in slots
Range:
1 to 1000
*RST:
20
<ACLRenable>
OFF | ON
Enable or disable the measurement of ACLR results
ON
: ACLR results are measured according to the other enable
flags in this command. ACLR results for which there is no
explicit enable flag are also measured (e.g. power in assigned
E-UTRA channel).
OFF
: No ACLR results at all are measured. The other enable
flags in this command are ignored.
*RST:
OFF
<UTRA1enable>
OFF | ON
Enable or disable evaluation of 1
st
adjacent UTRA channels
*RST:
OFF
<UTRA2enable>
OFF | ON
Enable or disable evaluation of 2
nd
adjacent UTRA channels
*RST:
OFF
<EUTRAenable>
OFF | ON
Enable or disable evaluation of 1
st
adjacent E-UTRA channels
*RST:
OFF
Example:
See
Firmware/Software:
V2.0.10
Options:
R&S CMW-KM012
CONFigure:LTE:MEAS<i>:MEValuation:LIST:SEGMent<no>:PMONitor
<Enable>
Enables or disables the measurement of power monitor results for segment <no>.
Suffix:
<no>
.
1..2000
Segment number
Parameters:
<Enable>
OFF | ON
*RST:
OFF
Example:
See
Firmware/Software:
V2.0.20
Options:
R&S CMW-KM012
Command Reference