LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
243
User Manual 1173.9628.02 ─ 16
2.5.3 RLC Throughput Tests
The RLC throughput measurement provided by the LTE signaling application is pro-
grammed as follows:
●
The application is controlled by SCPI commands with the following syn-
tax:
...:LTE:SIGN:THRoughput:...
●
After a
*RST
, the measurement is switched off.
You can start the measurement using
INIT:LTE:SIGN:THRoughput
and
retrieve the results using
FETCh:LTE:SIGN:THRoughput:...?
.
The examples in this section focus on commands directly related to the RLC through-
put measurement. For general configuration of the signaling application, refer to
ter 2.5.1, "General Configuration"
2.5.3.1
Configuring an RLC Throughput Measurement
// *****************************************************************************
// System-Reset
// *****************************************************************************
*RST; *OPC?
*CLS; *OPC?
// *****************************************************************************
// Configure repetition mode, update interval and window size.
// *****************************************************************************
CONFigure:LTE:SIGN:THRoughput:REPetition SINGleshot
CONFigure:LTE:SIGN:THRoughput:UPDate 200
CONFigure:LTE:SIGN:THRoughput:WINDow 20000
// *****************************************************************************
// Select the data application mode
// *****************************************************************************
CONFigure:LTE:SIGN:CONNection:CTYPe DAPPlication
2.5.3.2
Setting up a Data Connection
Proceed as follows:
1. Configure the other settings of the signaling application compatible to your UE and
configure the data application unit (see DAU documentation).
2. Switch on the cell signal and attach the UE.
3. Set up a data connection.
4. Generate IP traffic, for example using the IPerf measurement provided by the DAU.
Programming