![R&S CMW-KM5 Series Скачать руководство пользователя страница 230](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802230.webp)
LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
230
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:SIGN:MMONitor:IPADdress IP2
CONFigure:LTE:SIGN:MMONitor:IPADdress?
2.5.1.17
Attaching the UE and Activating an SCC
// *****************************************************************************
// Connect the UE (switched off). Switch on the DL signal. Query the cell
// state until it equals ON,ADJ (DL signal available at RF connector).
// *****************************************************************************
WAITKEY >Ensure that the UE is connected to the instrument and switched off<
SOURce:LTE:SIGN:CELL:STATe ON
WHILE SOURce:LTE:SIGN:CELL:STATe:ALL? <> "ON,ADJ"
// *****************************************************************************
// Switch on the UE and wait until it is attached (connection state = ATT).
// *****************************************************************************
WAITKEY >Switch on the UE<
WHILE FETCh:LTE:SIGN:PSWitched:STATe? <> "ATT"
// *****************************************************************************
// Query the RRC connection state.
// *****************************************************************************
SENSe:LTE:SIGN:RRCState?
// *****************************************************************************
// Query the SCC state. The SCC connection is set up automatically after attach
// if: a carrier aggregation scenario is active, the SCC activation mode is
// "AUTO" and keep RRC connection is enabled.
// *****************************************************************************
FETCh:LTE:SIGN:SCC:STATe?
2.5.1.18
Configuring the I/Q Settings
// *****************************************************************************
// Query the properties of the outgoing baseband signal, required to configure
// the baseband input of the connected instrument. Configure the baseband input
// according to the baseband output of the connected instrument.
// Configure also the SCC.
// *****************************************************************************
SENSe:LTE:SIGN:IQOut:PCC:PATH1?
SENSe:LTE:SIGN:IQOut:PCC:PATH2?
CONFigure:LTE:SIGN:IQIN:PCC:PATH1 -30, -20
CONFigure:LTE:SIGN:IQIN:PCC:PATH2 -30, -20
SENSe:LTE:SIGN:IQOut:SCC:PATH1?
SENSe:LTE:SIGN:IQOut:SCC:PATH2?
CONFigure:LTE:SIGN:IQIN:SCC:PATH1 -30, -20
CONFigure:LTE:SIGN:IQIN:SCC:PATH2 -30, -20