LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
156
User Manual 1173.9628.02 ─ 16
Fig. 2-67: UE identity settings
Default IMSI
15-digit International Mobile Subscriber Identity (IMSI). The default IMSI is required for
mobile terminated connection setups without previous attach of the UE.
Remote command:
2.4.11.6
Timer and Constants
The parameters in this section configure timers and counters.
Fig. 2-68: Timer settings
Out of Sync
The "Out of Sync" timer specifies the time after which the instrument, having waited for
a signal from the connected UE, releases the connection. The timer is started when an
uplink grant message is sent to the UE.
Remote command:
T3412
Timer "T3412" controls the initiation of a periodic tracking area update by the UE.
If the timer is disabled, no periodic tracking area update is required.
Remote command:
2.4.11.7
Time
The "Time" section allows you to send configurable date and time information to the
UE. Thus you can update the date and time displayed by the mobile. In a real network
this service is typically used to send the current local time to the UE.
The section is only visible if R&S CMW-KS510 is available.
GUI Reference