WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
98
User Manual 1173.9657.02 ─ 11
2.3.4 Maximum Power Measurements with E-DCH
This application sheet describes how to measure the maximum output power for a
WCDMA signal with HS-DPCCH and E-DCH, as defined in 3GPP TS 34.121, section
5.2B.
2.3.4.1
Options and Equipment Required
The test procedure described in this application sheet requires the following equip-
ment:
●
Wideband Radio Communication Tester R&S CMW500 or R&S CMW280 with soft-
ware version
≥
V3.0.30. The latest software version is recommended. This applica-
tion sheet describes software version V3.0.30.
The tester must be equipped with a wideband signaling unit.
●
Option R&S CMW-KS400, "WCDMA R99, basic signaling"
●
Option R&S CMW-KS401, "WCDMA R5/6, basic signaling"
●
Option R&S CMW-KS411, "WCDMA R5/6, advanced signaling"
●
Option R&S CMW-KM400, "WCDMA R99, TX measurement, uplink"
2.3.4.2
Test Overview
The conformance test specification 3GPP TS 34.121, section 5.2B "Maximum Output
Power with HS-DPCCH and E-DCH" defines a test for verification of the maximum UE
power with active HS-DPCCH and E-DCH. The test comprises five subtests with differ-
Application Sheets
深圳德标仪器