WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
595
User Manual 1173.9657.02 ─ 11
measurements that are essentially intended for checking limits, e.g. production
tests.
Remote command:
Measure on Exception
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected. A faulty result occurs e.g. when an overload is detected. In remote
control, the cause of the error is indicated by the "reliability indicator".
●
Off:
Faulty results are rejected. The measurement is continued; the statistical
counters are not re-set. Use this mode to ensure that a single faulty result does not
affect the entire measurement.
●
On:
Results are never rejected. Use this mode e.g. for development purposes, if
you want to analyze the reason for occasional wrong transmissions.
Remote command:
Measurement Length
Defines the number of consecutive slots that form a single measurement interval. The
measured slots are displayed in all multislot diagrams, e.g. "Error Vector Magnitude",
"CDP vs. Slot", and "UE Power".
See also
chapter 3.2.1.3, "Defining the Scope of the Measurement"
Remote command:
Preselected Slot
Selects the slot to be used for all single slot measurements,e.g. "Error Vector Magni-
tude vs Chip", "CD Monitor", "ACLR", "Emission Mask". The preselected slot must not
be confused with the slot used for tables of statistical values for multislot measure-
ments, see parameter
See also
chapter 3.2.1.3, "Defining the Scope of the Measurement"
Remote command:
Synchronization
Selects a slot number (0 to 14) that the R&S
CMW will display as the first slot in the
measurement interval. "Any" means that the measurement will start as fast as possi-
ble, beginning with the first captured slot. "Free Run" measurements use always "Any".
Selecting a synchronization slot number can speed up the synchronization process.
The trigger settings must be configured according to the selected slot. Example: To
use an external frame trigger with synchronization slot number 5, a trigger delay corre-
sponding to 5 slots has to be entered. Omitting the trigger delay results in a synchroni-
zation error because slot number 5 is expected but slot number 0 is found after the
measurement has been triggered.
Remote command:
GUI Reference