WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
583
User Manual 1173.9657.02 ─ 11
3.2.6.12
Using Markers
Use the "Marker" parameters to activate markers and to modify their position. The fol-
lowing "Marker" hotkeys are available at the bottom of the GUI:
Hotkey
Description
"Ref. Marker ..."
Enable or disable the reference marker, select a trace and the marker position
on that trace.
"Marker 1 /2 ..."
Enable or disable marker 1 or 2 and define the marker position (absolute or rela-
tive to the reference marker).
Depending on the trace mode, a trace can also be selected.
"Select Trace Mode"
Define whether all markers are collectively set to the same trace or to individual
traces.
See also: "Markers" in the R&S
CMW user manual, chapter "System Overview"
3.2.6.13
Common View Elements
Below the title bar, all views show the most important RF and analyzer settings as
shown below.
chapter 3.3.2.1, "Signal Routing and Analyzer Settings"
Tables
Most detailed views show tables providing a statistical evaluation of results measured
in a selected slot (or half-slot). For multislot measurements (e.g. "CDE vs Slot") the
"Slot Number Table" is used while for single slot measurements (e.g. "Emission Mask",
"EVM vs Chip") the "Preselected Slot" is used.
The selected slot number is displayed in most tables. For half-slot measurements also
the selected half-slot is displayed. Modify the (half-) slot to display the results of
another (half-) slot. A restart of the measurement is not required.
The statistical values in the tables are calculated as follows:
●
Current
: Value of the result obtained in the last measurement interval. For some
modulation results the current RMS value (the average over all samples in the
selected (half-) slot except the guard period) and the current peak value (the peak
of all samples in the selected (half-) slot except the guard period) are available.
●
Average
: Average of all "Current" values referenced to the last statistics cycle.
●
Max
: Largest or smallest "Current" value that the R&S
CMW obtained since the
start of the measurement.
●
StdDev
: Standard deviation of all "Current" values since the start of the measure-
ment.
All statistical results (statistical tables and "Average" or "Max" traces) are calculated
according to the general rules for statistical results.