WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
174
User Manual 1173.9657.02 ─ 11
When one of the buttons is pressed for TPC setup "Max. Power E-DCH", an initial
check is performed before the action is executed. The check verifies that both the tar-
get E-TFCI and the current E-TFCI can be determined and are equal. If this is not the
case, the initiated action is aborted and the "TPC Condition" indicates the error (e.g.
"Missing Resource" or "Setting Conflict").
Remote command:
TPC Condition
Displays the current TPC state. Transition states that would be displayed for a very
short time only are indicated via remote command, but not displayed at the GUI (e.g.
transmission of single pattern).
Possible values are:
●
Idle:
no connection established
●
Continuous Pattern:
transmitting continuous pattern
●
Alternating:
transmitting alternating pattern
●
Prec. <Precondition> (press Execute):
indicated <Precondition> has been
reached
●
<State> (press Precond. or Execute):
The current <State> results from a previ-
ously executed TPC setup and does not match the precondition of the active TPC
setup.
●
Target Power Locked:
closed loop target power reached
●
Target Power Unlocked:
reaching closed loop target power failed
●
Max Power:
maximum power reached
●
Min Power:
minimum power reached
Values only relevant for "Max. Power E-DCH" setup:
●
Missing Resource:
required resources are in use by another measurement
●
Searching:
setup started, max power not yet reached
●
Failed:
test procedure failed in state "Searching", see "Max. Power E-DCH Condi-
tion" for details
●
Setting Conflict:
settings inappropriate for the setup, see "Max. Power E-DCH
Condition" for details
●
Settings Changed:
relevant settings changed after setup execution
Remote command:
Max. Power E-DCH Condition
This parameter is visible if the TPC setup "Max. Power E-DCH" is active (requires
option R&S CMW-KS401).
It displays two E-TFCI values:
●
Target E-TFCI: expected E-TFCI value, calculated from the HSUPA settings
●
Current E-TFCI: value sent by the UE, monitored repeatedly during execution of
the setup, each time for 150 ms
If several E-TFCI values have been monitored within the 150 ms, the smallest
value is displayed. Exception: If the smallest value equals the target E-TFCI, the
next larger monitored value is displayed instead.
GUI Reference