Remote control commands
R&S
®
NRPxxP
59
User Manual 1179.5760.02 ─ 02
8.6.2.1
Defining the dropout tolerance
To prevent power drops due to modulation from being erroneously interpreted as the
end of a pulse, you must define the dropout tolerance. The dropout tolerance is a time
interval in which the pulse end is only recognized if the signal level no longer exceeds
the trigger level.
SENSe<Sensor>:[POWer:]BURSt:DTOLerance
<tolerance>
Sets the dropout tolerance, a time interval in which the pulse end is only recognized if
the signal level no longer exceeds the trigger level. See
.
Parameters:
<tolerance>
Range:
0.00 to 0.003
*RST:
1.000e-6
Default unit: Seconds
8.6.2.2
Defining a time interval for the measurement
At the beginning and at the end of the measurement interval, you can define time inter-
vals that are excluded from the measurement, see
Chapter 8.7.3, "Excluding intervals"
8.6.2.3
Triggering a burst average measurement
In burst average mode, only internal trigger events from the signal are evaluated, irre-
spective of the setting of the
parameter. The
parameter is also ignored, so that the measurement interval begins exactly when the
signal exceeds the trigger level.
8.6.2.4
Querying the pulse interval
SENSe<Sensor>:[POWer:]BURSt:LENGth?
Queries the length of a burst (pulse interval), the time between the trigger point of the
measurement and the time the trigger logic detects the end of the pulse. See
Usage:
Query only
8.6.3
Timeslot measurement
The timeslot mode is used to measure the average power of a definable number of
successive timeslots within a frame structure with equal spacing. The measurement
result is an array with the same number of elements as timeslots. Each element repre-
sents the average power in a particular timeslot.
Configuring the measurement modes