Remote control commands
R&S
®
NRPxxP
48
User Manual 1179.5760.02 ─ 02
Example: Average count = 1
1
For average count 1, the setting of the termination control has no impact. In both
cases, the measurement runs for the duration of one aperture time. Then, settled data
is available, and the power sensor returns to the idle state.
1
2
3
4
5
1 = Trigger event
2 = Noninverted chopper phase
3 = Measurement result
4 = Trigger synchronization
5 = Return to idle state
8.5.4.2
Trace mode
General settings for the first two examples:
●
ON
●
SENSe<Sensor>:TRACe:AVERage:COUNt
2
●
SENSe<Sensor>:TRACe:AVERage[:STATe]
ON
Example: Repeating termination control
Further settings for this example:
●
SENSe<Sensor>:TRACe:AVERage:TCONtrol
REPeat
Every chopper phase is started by a trigger event and lasts the defined trace time. Dur-
ing a chopper phase, the trigger synchronization is high (
ON
).
After 2 chopper phases, 1 measurement is completed. As defined by the trace average
count, after 2 measurements, the trace measurement result is averaged and available.
1
2
3
4
5
7
6
1 = Start of the measurement cycle
2 = Trigger events
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trace measurement result
6 = Trigger synchronization
7 = Return to the start of the measurement cycle
Controlling the measurement