Remote control commands
R&S
®
NRPxxP
101
User Manual 1179.5760.02 ─ 02
Parameters:
<time>
Specified in percent of the pulse amplitude.
Range:
Depends on the trigger delay.
*RST:
0.0
Default unit: s
SENSe<Sensor>:TRACe:MEASurement:PULSe:DCYCle?
Queries the duty cycle of the measured power. Unit: %
Duty cycle =
Pulse period
Pulse width
100 % x
If the pulse period, and thus the duty cycle, cannot be determined, NaN for "not a num-
ber" is returned.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:PULSe:DURation?
Queries the time between the first positive edge and the subsequent negative edge of
the pulse, where the edges occur at crossings of the mid threshold.
If the pulse period, and thus the duty cycle, cannot be determined, NaN for "not a num-
ber" is returned.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:PULSe:PERiod?
Queries the time between two consecutive edges of the same polarity in seconds. Dur-
ing this time, the pulse signal completes one cycle.
If the trace covers at least three complete periods, the power sensor automatically
detects double pulses and determines their period.
If the pulse period, and thus the duty cycle, cannot be determined, NaN for "not a num-
ber" is returned.
Usage:
Query only
Querying measurement results