Measurement Types and Result Displays
R&S
®
NRX
85
User Manual 1178.5566.02 ─ 07
Pulse Top ← Power tab
Pulse top power level detected by the selected
. This value is used as a refer-
ence (100
%) to determine other parameter values such as the rising or falling thresh-
olds.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:PULSe:TOP[:
STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:PULSe:TOP?
Trace Peak ← Power tab
Maximum power measured within the analysis window.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:MAXimum[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:MAX?
Neg. Overshoot ← Power tab
Height of the local minimum before a rising edge, divided by the pulse amplitude:
Negative overshoot = 100 % x
Pulse base power - minimum power
Pulse amplitude
Depends on the setting under
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:NEGative:
OVERshoot[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:NEGative:
OVERshoot?
Pulse Analysis
www.allice.de
Allice Messtechnik GmbH