Remote Control Commands
R&S
®
NRX
374
User Manual 1178.5566.02 ─ 07
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
1.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio:RFRatio
<value>
is set to:
REC
,
FREC
or
RREC
Sets the ratio of forward/reverse power that corresponds to 2.5
V output voltage.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
100.0
Default unit: pct
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio:RLOSs
<value>
is set to:
REC
,
FREC
or
RREC
Sets the return loss measurement value that corresponds to 2.5
V output voltage.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to 180.0
*RST:
10.0
Default unit: dB
Manual operation:
See
Configuring the Analog Signal Ouput and the Trigger Input/Output