Remote Control Commands
R&S
®
NRX
372
User Manual 1178.5566.02 ─ 07
Sets the standing wave ration (SWR) measurement value that corresponds to 0
V out-
put voltage.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
1.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio[:VALue]
<value>
is set to:
REC
,
FREC
or
RREC
Sets the power ratio measurement value that corresponds to 0
V output voltage.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +180.0
*RST:
+0.0
Default unit: dB
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:CCDF
<value>
is set to:
REC
,
FREC
or
RREC
Sets the CCDF measurement value that corresponds to 2.5
V output voltage.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
1.0
Default unit: pct
Configuring the Analog Signal Ouput and the Trigger Input/Output