Remote Control Commands
R&S
®
NRX
322
User Manual 1178.5566.02 ─ 07
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<duty_cycle>
Range:
0.001 to 100.0
*RST:
50.0
Default unit: pct
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:ESAMpling
<value>
Effective for trace measurements.
Enables or disables the automatic equivalent sampling that allows for high resolution
measurements.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:SAMPling
<value>
Effective for continuous average measurements.
Sets the sampling rate.
If the sampling frequency is located within the video bandwidth, aliasing can occur, that
is when spectral components near the sampling frequency cause beating effects. If you
change the sampling rate, the beating effects usually disappear.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
FREQ1 | FREQ2
FREQ1
Normal sampling rate
Configuring Sensors