Remote Control Commands
R&S
®
NRX
278
User Manual 1178.5566.02 ─ 07
Usage:
Query only
CALCulate<Measurement>:STATistics:MARKer:X:POSition[:ABSolute]
<value>
Effective for statistics measurements.
Sets the absolute position of the x-marker.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dBm
Manual operation:
See
CALCulate<Measurement>:STATistics:MARKer:X:POSition:RELative
<value>
Effective for statistics measurements.
Sets the x-marker relative to a power value.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>:STATistics:PDF:MARKer:Y:POSition
<value>
Positions the y-marker to a measurement value of the PDF statistics measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 10000.0
*RST:
0.0
Default unit: -
Manual operation:
See
Measurement Settings and Results