Remote Control Commands
R&S
®
NRX
256
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence?
Effective for pulse analysis measurements.
Queries the time when the signal passes through the medial reference power level with
rising edge, referenced to the delayed trigger event. Indicates the start point of the first
power pulse within the analysis window.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OVERshoot?
Effective for pulse analysis measurements.
Queries the height of the local maximum before a falling edge, divided by the pulse
amplitude:
Positive overshoot = 100 % x
Max. power - pulse top power
Pulse amplitude
Depends on the setting under
DISPlay[:WINDow<Window>]:TRACe:
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:SPERiod?
Effective for pulse analysis measurements.
Queries the number of samples per second.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation: