Remote Control Commands
R&S
®
NRX
250
User Manual 1178.5566.02 ─ 07
Parameters:
<time>
Range:
0.0 to 30.0
*RST:
0.01
Default unit: s
CALCulate<Measurement>:TRACe:MEASurement:ALGorithm
<value>
Effective for pulse analysis measurements.
Sets the analysis algorithm for detecting the pulse top and the pulse base power of a
pulsed signal. From these two power levels, the reference levels are derived.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
HISTogram | INTegration | PEAK
*RST:
HISTogram
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:DEFine:DURation:REFerence
<value>
Effective for pulse analysis measurements.
Sets the pulse width, pulse start time and pulse stop time.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
50.0
Default unit: pct
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:DEFine:TRANsition:
HREFerence
<value>
Effective for pulse analysis measurements.
Defines the end of the rising edge and the start of the falling edge of the pulse. Used
for the measurement of the rise or fall time.
Suffix:
<Measurement>
.
1 to 8
Measurement
Measurement Settings and Results