Remote Control Commands
R&S
®
NRX
231
User Manual 1178.5566.02 ─ 07
Manual operation:
CALCulate<Measurement>:TYPE
<type>
Sets the measurement type.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<type>
CONTav | NRT | TRACe | STATistics | TGATe | BURStav |
TSLot | PULSe
*RST:
CONTav
Manual operation:
[SENSe<Sensor>:]AUXiliary
<mode>
Enables the measurement of additional measured values that are determined together
with the main measured value.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<mode>
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
Minima and maxima of the trace are transmitted together with
the measured value.
Usually, extreme values are lost due to averaging the measured
values.
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH