Firmware Update
R&S
®
NRX
168
User Manual 1178.5566.02 ─ 07
3. Register the instrument as a VISA device. Refer to documentation of your VISA
software for details.
12.1.3
Updating the Application Firmware
To perform a firmware update:
1. Start the Firmware Update program (PureFW) via "Start menu > NRP-Toolkit >
Firmware Update". The following window should appear:
The program automatically starts scanning for R&S power sensors and meters
attached via USB. When the scan is completed, all recognized power sensors and
meters are listed in the "Device" dropdown control.
2. If the instrument you want to update is not listed in the "Device" dropdown control,
perform one of the following:
a) If the instrument is connected to the PC via USB, press "Rescan" to search for
R&S power sensors and meters attached via USB.
Firmware Update via PC and USB or Ethernet Connection