Remote Control Commands
R&S
®
NRP18S-xx
113
User Manual 1178.3686.02 ─ 05
Use this query to determine a useful resolution for the result display near the lower
measurement limit.
Usage:
Query only
SYSTem:PARameters?
Lists all commands with default values, limits and ranges.
Usage:
Query only
SYSTem:PARameters:DELTa?
Lists all commands that differ from the defined default status set by
The commands are output with default values, limits and ranges.
Usage:
Query only
SYSTem:PRESet
Triggers a sensor reset.
The command essentially corresponds to the
*RST
command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
SENSe:TRACe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the sensor.
Usage:
Event
SYSTem:RESTart
Restarts the firmware of the R&S NRP18S-xx.
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode.
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(
ON).
Configuring the System